SensoLite Nova
4. The meter returns automatically to the test screen where the recently set code value is
displayed, showing that it has been stored in SensoLite Nova. Now the meter is ready to
perform the blood glucose test with the new code value.
Warning!
SensoLite Nova waits for the insertion of the Code-card for 60 seconds. If the Code-card is
not inserted within 60 seconds the meter displays an E-C error message and the actual code
value remains the same as before. In this case please repeat the coding process again.
Ready for measurement!